M210A1

Amplifier

The M210А1 amplifier provides:

– verification and calibration of low-power voltage transformers (LPVT) as part of any VT calibration system;
– conversion of any secondary voltage of a traditional VT into a value equal to the nominal voltage of the reference VT.

FEATURES AND BENEFITS

  • Continuous range of rated secondary voltages of verified LPVT
  • Continuous range of rated secondary voltages of verified traditional VT
  • High conversion accuracy
  • Simple use by intuitive user interface
  • Storage of individual voltage settings
  • Possibility of using with automated VT calibration system

 


APPLICATIONS

M210A1 is used by:

  • Manufacturers of LPVT
  • Calibration Laboratories
  • Metrology Institutes
Input 1:
Rated input voltage range (Uin1r)0.1 V … 4 V
Input voltage resolution0.001 V
Additional factors× 1/√3; × 1/3
Rated operating frequency50/60 Hz
Input impedance, represented as a parallel resistor-capacitor circuit2 МΩ ± 5% || C ≤ 50 pF
Input 2:
Rated input voltage range (Uin2r)30 V … 245 V
Input voltage resolution0.1 V
Additional factors× 1/√3; × 1/3
Rated operating frequency50/60 Hz
Input impedance, represented as a parallel resistor-capacitor circuit2 МΩ ± 5% || C ≤ 50 pF
Output:
Rated output voltage range (Uoutr)30 V … 100 V
Output voltage range0 … 120 V
Output voltage resolution0.1 V
Additional factors× 1/√3; × 1/3
External burdenR ≥ 500 kΩ; C ≤ 5 nF
Accuracy:
Ratio error
0.02·Uoutr ≤ Uoutout < 0.2·Uoutr
0.2·Uoutr ≤ Uout < 0.8·Uoutr
0.8·Uoutr ≤ Uout < 1.9·Uoutr

± 0.1 %
± 0.04 %
± 0.02 %
Phase displacement
0.02·Uoutr ≤ Uoutout < 0.2·Uoutr
0.2·Uoutr ≤ Uout < 0.8·Uoutr
0.8·Uoutr ≤ Uout < 1.9·Uoutr

± 5 min (± 0.15 crad)
± 2 min (± 0.06 crad)
± 1 min (± 0.03 crad)
Power mains:
Voltage
Rated frequency
100…240 V
50/60 Hz
Operating temperature
Relative humidity
0...40 °C
up to 80 % non-condensing
Size
Weight
245 × 135 × 325 mm
5 kg

M210A1 test arrangement 1

CL – low-voltage measuring capacitor; CH – high-voltage measuring capacitor; CC – current comparator with automatic balancing

Fig.1 Test circuit for LPVT by use the Voltage Transformer Test Set M210/M211

M201A1 test arrangement 2

VTN – standard transformer
Fig.2 Test circuit for LPVT by use the Comparator M100

M210A1 test arrangement 3

VTN – standard transformer; VTX– transformer under calibration
Fig.3 Test circuit for traditional voltage transformers with different rated secondary voltages

RELATED PRODUCTS

M210/M211

Voltage Transformer Test Set

M410

Electronic Voltage Burden

M411

Electronic Voltage Burden