M400

Electronic Current Burden (1A)

 

Electronic Current Burden M400 is designed to provide a programmable burden in the secondary circuits of current transformers during verification, calibration, and testing. The device is intended for use with current transformers having a nominal secondary current of 1 A

FEATURES AND BENEFITS

  • Wide range of burden values according to standards: IEC 61869-2, IEC 60044-1, IEEE C57.13
  • Rated current – 1 A
  • Rated power range: 0,5 to 50 VA, cos φ = 0,8 at 50 Hz 0; 0,5 to 50 VA, cos φ = 1 at 50 Hz
  • Input cable, contact and internal test set resistances are compensated by four-conductor connection
  • Non-breaking measuring circuit during switching of the burden
  • Possibility of using with automated current transformer calibration systems
  • Small size and low weight

 

VERSIONS

Electronic Current Burden М400 is available in two versions:

  • М400.1 allows current transformer calibration in secondary current range 1…120 % of rated current
  • М400.2 allows current transformer calibration in secondary current range 1…150 % of rated current

 

APPLICATIONS

Electronic Current Burden M400 is used by:

  • Manufacturers of Current Transformers
  • On-site testing of Current Transformers
  • Calibration laboratories
  • Metrology institutes

Specification

Value

Rated current

1 A

M400.1 / M400.2

Current range

1…120 %

M400.1
1…150 %

M400.2

Rated burden power for cos φ = 0,8 0,5; 0,75; 1; 1,25; 1,5; 1,75; 1,875; 2; 2,5; 3; 3,75; 4; 5; 6,25; 7,5; 10; 11,25; 12,5; 15; 20; 25; 30; 40; 45*; 50* VA

M400.1 / M400.2

Rated burden power for cos φ = 1 0; 0,5; 0,75; 0,8; 1; 1,25; 1,5; 1,875; 2; 2,5; 3,75; 4; 5; 6; 7,5; 10; 12,5; 15; 30; 50* VA

M400.1 / M400.2

Active and reactive power error

± 0,03·S**  for 0,8 < S < 50 VA

M400.1 / M400.2

± 0,04·S  for S = 0,75 VA

± 0,06·S  for S = 0,5 VA

± 0,05 VA for S = 0 VA

* Burden power values are possible for 1 to 120 % of rated current.

** Setted burden power value.

 

 

 

Power mains

Rated voltage

Rated freguency

220/230 V 50 Hz
Operating temperature 0…40 °С
Relative humidity up to 80 % non-condensing
Size (W × H × D) 250 × 140 × 330 mm
Weight (Main unit) 5 kg
Standards

Safety

EMC

  EN 61010-1:2010 EN 61326-1:2013

*** The manufacturer reserves the right to change the characteristics of the device.

RELATED PRODUCTS

S100

Automatic Сurrent  Source

M401

Electronic Current Burden (5A)

M402

Electronic Current Burden (1; 2; 5 A)

M500

Programmable Standard Current Transformer

M100

Comparator